Signature® Flow Meters

The Signature® flow meter is designed for open channel flow monitoring applications. It supports flow measurement technologies including ultrasonic, bubbler, submerged ultrasonic area velocity, and non-contacting laser Doppler area velocity. The meter can calculate flow using standard open channel level-to-flow conversions, as well as user defined equations or level to flow data points, depending on the application need. 

With multiple smart interface options and multi-parameter logging (such as pH), the Signature flow meter provides a common platform for control action, reporting, and communication. 

 

     Flexible, configurable flow meter for permanent discharge compliance monitoring


 

 Signature® Flow Meter

Flexible, configurable flow meter for permanent discharge compliance monitoring 

The Signature flow meter is designed for open channel flow monitoring applications. It supports flow measurement technologies including bubbler, non-contact laser area velocity, submerged Doppler ultrasonic area velocity, and ultrasonic. The meter can calculate flow using standard open channel level-to-flow and area velocity conversions, as well as user-defined equations, level to area data points, or level to flow data points, depending on the application need. 

The Signature flow meter has unique features to verify data integrity. It logs key events such as changes in calibration and power outages to validate data accuracy. Data can be easily reviewed to detect any type of data alteration. With multiple smart interface options and multi-parameter logging (such as pH), the Signature flow meter provides a common platform for control action, reporting, and communication. 

Multiple flow technologies are available so you can select the best flow technology for the measurement site and channel conditions. Available technologies are: bubbler, ultrasonic, continuous wave area velocity, and non-contacting LaserFlow™ velocity.

Standard Features

  • Multiple simultaneous flow technologies 
  • SDI-12 input 
  • RS-485 Modbus input 
  • RS-485 Modbus output 
  • NEW! Rain gauge input 
  • Multiple parameter data logging 
  • Program and Summary Reports 
  • Data integrity verification 
  • Triggering, sampler enabling 
  • Compatibility with Flowlink® software

Applications

  • Industrial Pretreatment Compliance 
  • Shallow Flow Measurement in Large and Small Pipes 
  • Permit Enforcement 
  • Wastewater Treatment Plants 
  • Outfall

Options and Accessories

  • TIENet® 301 pH/Temperature Device 
  • NEW! TIENet 304 Contact Output Card (two user-assignable outputs per card) 
  • TIENet 306 Sampler interface (triggering and pacing) 
  • NEW! TIENet 307 Analog Input Card (two user-assignable inputs per card) 
  • TIENet 308 Analog Output Card (two user-assignable outputs per card) 
  • TIENet 310 Ultrasonic Level Sensor 
  • TIENet 330 Bubbler module 
  • TIENet 350 Area Velocity Sensor 
  • TIENet 360 LaserFlow™ Non-contacting Area Velocity Sensor 
  • Battery back-up power 
  • Mechanical totalizer 
  • Advanced cellular and Ethernet communication options 
  • Mounting hardware
 

 

Specifications

Signature® Flow Meter

Size (HxWxD): 8.88 x 12.22 x 8.22 in. (with mounting bracket), 16.74 x 13.58 x 10.48 in (with stand)
Materials: PPO Polyphenylene Oxide
Enclosure (self-certified): NEMA 4X/IP66
Power Required: 100 to 240 VAC, 50/60 Hz
Cable Entry: Standard- ¾" NPT conduit, optional- ¾" NPT cord grips
Flow Measurement Technologies: Ultrasonic (TIENet 310), Bubbler (TIENet 330), Area Velocity (TIENet 350, 360)
Inputs: Two SDI-12, two MODBUS ASCII/RTU, pH Measurement (TIENet 301), Analog In (TIENet 307), Rain In
Setup: Front Panel Keypad, Flowlink Software - with serial USB, remote cellular, or Ethernet
Flow Conversions: Area Velocity, Weir, Flume, British Flume, Metering Insert, Manning Formula, Equation, Level to Flow Data Points, Level to Area Data Points
Data Storage: Non-volatile flash; retains stored data during program updates. Interval- 15 or 30 seconds; 1, 2, 5, 15, or 30 minutes; or 1, 2, 4, 12, or 24 hours. Capacity- 8 MB (180 days with 5 parameters logged at 1 minute intervals, reports once per day).
Data Retrieval: USB drive, Flowlink Software - with serial USB, remote cellular, or Ethernet
Outputs: MODBUS ASCII/RTU, Analog (TIENet 308), Contact Output (TIENet 304), SMS Alarm
Sampler Interface: TIENet 306

TIENet® 301 pH/Temperature Device

Weight (w/o probe): 3.5 lb w/ 10 m cable, 7.5 lb w/ 23 m cable
Ambient Operating Temperature: -20 to 50 °C (-4 to 122 °F)
pH Measurement Range: 0 - 14 pH units
Temperature Compensation: Performed by the 301 device
pH Accuracy: ±0.1 pH units (new probe, freshly calibrated w/in range)
Probe Dimensions: 1.12" X 6" long, 3/4 NPT; Cable 25 ft
Probe Body Material: 316SST
pH Electrode Junction: Double porous
Temperature Measurement Range: 0 to 80 °C (32 to 176 °F)

TIENet® Model 304 Contact Output

Type: NO or NC, Galvanically isolated
Output rating: 30 volts, 1 Ampere
Operating Temperature: -20 to 70 °C (-4 to 158 °F)

TIENet® Model 306 Sampler Interface

Function: Flow pacing, enabling based on triggered event. Time and bottle information sent to Signature Flow Meter
Powered By: Signature Flow Meter
Operating Temperature: -20 to 50 °C (-4 to 122 °F)
Storage Temperature: -40 to 60 °C (-40 to 140 °F)
Pulse Width: 50 ms
Pulse Output: 5 volts
Sampler Connection: Standard 6 pin connector for Isco 6712, Avalanche, Glacier, GLS, and 3700 samplers. For other options, contact factory

TIENet® Model 307 Analog Input

Power source in active mode: 17-29 volts
Isolation: Galvanically isolated
Input Impedance: 400 ohm
Operating Temperature: -20 to 70 °C (-4 to 158 °F)

TIENet® Model 308 Analog Output

Output: 4-20 mA
Isolation: Monolithic
Maximum Load: 900 ohm
Outputs per Card: Two

TIENet® Model 310 Ultrasonic Level Sensor

Measurement Range: 0.3 to 3.3 m (1 to 11 ft)
Accuracy at 72 °F (22 °C): ±0.006 m (0.02 ft) at less than or equal to 1 ft level change, ±0.009 m (0.03 ft) at greater than 1 ft level change
Temperature Coefficient (w/ in compensated range): ±0.0002 x Distance (m) x Temperature Deviation from 22 °C. ±0.00011 x Distance (ft) x Temperature Deviation from 72 °F.
Beam Angle: 10° (5° from center line)
Frequency: 50 kHz
Size: 9.1 cm diameter X 10.2 cm tall (3.63"x 4")
Cable Length: 10 or 23 m (32.8 or 75.5 ft)
Weight: 1.8 kg (4 lbs)
Body Material: PVDF
Temperature Range: -30° to 60 °C, -22° to 140 °F (Operating & Storage)
Certifications: Group 2, Category 1G (zone 0), T4, Class I, Div 1, Groups C & D, T4 (pending)

TIENet® 330 Bubbler Module

Level Measurement Range: 0.003 to 3.05m (0.01 to 10 ft)
Level Measurement Accuracy: +0.002m @ 22 °C (0.007 ft @ 72 °F)
Operating and Storage Temperature: -18° to 60 °C (0 to 140 °F)
Temperature Compensation Range: 0° to 60 °C (32° to 140 °F)
Temperature Coefficient (w/in compensated range): ±0.0003 x Level (m) x Temperature Deviation from 22 °C. ±0.00017 x Level (ft) x Temperature Deviation from 72 °F.

TIENet® Model 350 Area Velocity Sensor

Probe Dimensions: 1.9 x 3.3 x 15.2 cm (0.75 x 1.3 x 6.0 in)
Materials: Sensor- epoxy, chlorinated PC, SST; Cable- UV-Rated PVC
Temperature Range: 0 to 70 °C (32 to 158 °F)
Velocity Measurement Range: -1.5 to 6.1 m/s (-5 to 20 ft/s)
Velocity Measurement: Bi-Directional
Velocity Accuracy: ±0.03 m/s (±0.1 ft/s) from -5 to 5 ft/s; ±2% of reading from 5 to 20 ft/s, Uniform velocity profile
Minimum Depth: 25 mm (0.08 ft)
Frequency: 500 kHz
Level Measurement Range: 0.01 to 3.05 m (0.033 to 10 ft)
Level Accuracy: ± 0.10% Full Scale
Maximum Allowable Depth: 10.5 m (34 ft)
Typical Long-Term Stability: ±0.007 m/yr (±0.023 ft/yr)

TIENet® Model 360 LaserFlow Velocity Sensor

Sensor Dimensions: 38.01 x 26.21 x 56.7 cm, 14.96 x 10.3 x 22.32 in
Weight: 8.7 kg (19.2 lbs)
Materials: Conductive Carbon Filled ABS, SST, Conductive Kynar, Anodized Aluminum, UV-Rated PVC
Temperature Range: Operating 0 to 60 °C (32 to 140 °F). Storage -40 to 60 °C (-40 to 140 °F)
Cable Lengths: 5, 10, or 15 m (16.4, 32.8, or 49.2 ft)
Enclosure: IP68
Certifications: CE EN61326; FDA CDRH 21CFR1040; IEC 60825-1
Power Required: 8 to 26 VDC Input voltage, 12 VDC Nominal
Flow Accuracy: ±5% of Reading. (Typical, under normal flow conditions)
Velocity Measurement Range: -4.6 m/s to 4.6 m/s (-15 ft/s to 15 ft/s)
Direction: Bi-Directional
Velocity Accuracy: ±0.5% of reading ±0.03 m/s (0.1 ft/s)
Minimum Velocity: 0.25 m/s (0.8 ft/s)
Level Measurement Range: 0 to 3 m (0 to 10 ft) from measurement point
Level Accuracy @ 22 °C (72 °F): ±0.006 m (0.02 ft) at less than or equal to 1 ft level change; ±0.012 m (0.04 ft) at greater than 1 ft level change
Temperature Coefficient within compensated range: ± 0.0002 x D (m) per degree C; ± 0.00011 x D (ft) per degree F (Where D = Distance from transducer to liquid surface)
Beam Angle: 10° (5° from center line)
Ultrasonic Signal:

50 KHz