Laboratory and Industrial Gas Standards Generators

These instruments use permeation tubes to deliver gas standards directly to an FTIR, GC, GC-MS, Ion Mobility Spectrometer or any type of gas analyzer or environmental chamber.


 

FlexStream™ Gas Standards Generator

       The FlexStream™ gas standards generator provides:
Permeation tube oven with wide temperature range controlled to 0.01 °C
High-accuracy primary dilution mass flow rate
Front panel color touch-screen human interface
Automation of all calculations; key in the target concentration and the FlexStream sets the flow rate
Single-oven module can create mixtures with up to 8 components
External computer monitoring and control.

FlexStream features: 

  • Complete, integrated, ready-to-use turnkey system
  • Automated operation with direct readout of concentration in units of ppm, ppb or ppt
  • Modes of Operation: standby, zero, span and purge
  • Touch screen interface for local setup and front panel control
  • Modbus connectivity for integration with other computer systems for local or remote control
  • Designed for expandability 
  • Flow path designed for maximum flexibility with minimum error
  • Flow path suitable for reactive gases - mixture contacts only glass, Teflon® and stainless steel (other materials available) 
  • Accepts disposable permeation tubes, diffusion tubes, ultra-high rate liquid filled tubes, wafer tubes, and prefilled gas fed permeation tubes
  • Accepts up to 8 disposable tubes per oven with maximum 6 inch length x 1/4 inch diameter (KIN-TEK HRT, SRT and EL tubes)
  • Accepts one of KIN-TEK refillable LFH, ULED and 57 Series tubes per oven

 

Principle of Operation FlexStream
Base Module
     Dilution of constant component vapor emission from permeation or 
     diffusion tubes
     Delivery of precise, adjustable, low concentration dynamically blended
     gas mixtures. Trace concentrations - ppmv, ppbv, pptv
     Front panel touchscreen for oven temperature and dilution flow rate
     adjustment and standby, zero and span selection
     External PC monitoring and control via industry-standard
     Modbus® RTU communications protocol and Ethernet or serial
     RS-232 connectivity
   
Permeation Tube Oven  
     Construction - high thermal mass oven with electronic PID control Vertical orientation
     Oven temperature range (heat only) 30-150 °C
     NIST-traceable oven temperature
     Oven setpoint resolution and control ±0.01 °C
     Temperature digital indication and adjust via front panel
     Temperature digital indication and adjust via computer interface
     Over-temperature detection with audible alarm and oven power removal
   
Diluent Gas (customer-supplied)  
     Uses inert, non-reactive, non-hazardous dry gas (Nitrogen, zero air,
     Argon, etc.)
     Mass flow controller subsystem is factory-calibrated in Nitrogen; contact 
     factory for calibration in other type gas
     FlexSteam accepts a gas correction factor (GCF) for most gases to auto-adjust
     the dilution mass flow rate when a diluent gas other than Nitrogen is used
     Using the supplied FlexLink™ PC-based software, the mass flow controller
     subsystem can be easily calibrated with up to 10 points in other diluent gases
     (Customer supplies the flow standard)
       
Primary Dilution Mass Flow Controller  
     Standard mass flow controller range 0.25 to 5 lpm
     Optional mass flow controller ranges available (0.5-10, 0.1-1, 0.1-0.5 lpm)
     Mass flow rate accuracy (with multi-point calibration) Lesser of±1.5% Reading and ±1% FS
     Mass flow rate digital indication and control via front panel
     Mass flow rate digital indication and control via computer interface
     Mass flow controller calibration method: piece-wise linear fit up to 10 calibration points
   
Accepts Vapor Emissions from the Following Devices*  
     Liquid-filled disposable tubes up to 15 cm in length (maximum of 8 tubes)
     Liquid-filled high rate (LFH) permeation tubes (maximum of 1 tube)
     Factory pre-filled gas-fed permeation tubes (maximum of 1 tube)
     Diffusion tubes (maximum of 4 tubes)
   
*Note: A single oven can hold only one of the above tube types at a time  
   
Standard Component Flow Path  
     Component contacts only glass, Teflon® and stainless steel
     Glass permeation tube holder
     Suitable for reactive component gases
   
Optional Component Flow Path (specified at time of order)  
     Heated output lines
     Stainless steel
     Electro-polished stainless steel
     Sulfinert-coated stainless steel
     Teflon only
     Teflon permeation tube holder
   
Standard Modes of Operation  
     Standby (oven venting at approx 75 cc/min flow rate)
     Zero (with user-specified flow rate)
     Span-by-flow (with user-specified flow rate or concentration)
     Span-by-concentration (with user-specified target concentration)
     Purge
     Auto-calculates flow setpoint based on target concentration, 
     tube emission rate and oven temperature
     Automatically updates touchscreen with up to 8 generated concentration values
   
Output Concentration Range  
     Below 1 ppb to over 1000 ppm depending on tube emission rate 
     and dilution flow rate; low ppb and ppt concentrations with secondary dilution
     User selects units of ppm, ppb or ppt
   
Interfaces  
     Local interface - color touch screen display with virtual keypad
     Remote interfaces - Ethernet (standard) and serial RS-232 (standard)
     RS-232 conversion to RS-485 via customer-supplied converter (optional)
     Communications Protocol Modbus RTU
   
PC-based Software  
     KIN-TEK FlexLink™ software CD shipped with FlexStream Base module
     FlexStream Labview® Modbus driver on FlexLink™ CD
     LabView®-based starter software with source provided upon request
   
Expandability  
     Additional permeation ovens (via up to 5 FlexPM modules)
     Customer-refillable gas-fed permeation oven option (via FlexGF modules)
     Secondary diluition flow rate option (via FlexSD module); other ranges available

0.5 to 5 lpm

     Overall output gas stream dilution ratio with both primary and secondary dilution Up to 10,000 to 1
     Humidified gas module option (contact factory for more information)
     Gas stream output Interface module option (via FlexIM module);
     If FlexSD secondary dilution module is installed, subtract 10 psig (69 kPa)
ambient to 30 psig (207 kPa)
   
Instrument Housing (specified when ordering)  
     Carrying case (standard)
     Expansion chassis (option); desktop or rack-mount; 2 cases maximum 3 modules per case
     Maximum number of modules in a single FlexStream system 6
   
Power Requirements (specified when ordering)  
     115 VAC, 50/60 Hz, 2 A
     230 VAC, 50/60 Hz, 1 A
   
Dimensions and Weight  
     Dimensions (Base module as packaged in carrying case); approximate;
     add 3.5 in (8.3 cm) to Depth for front panel inlet filter clearance
7.5 W x 13.5 H x 20 D inch
18.4 W x 34.3 H x 50.8 D cm
     Weight (Base module as packaged in carrying case); approximate 27 lb (12.2 kg)
   
Certifications  
     European CE Mark (230 VAC, 50 Hz): FlexBase, FlexPM, FlexSD
     NRTL (site-specific)
     CSA (site-specific)
   
Speciality Applications (Please contact the Factory)  
     Special diluent (carrier) gases other than inert gases
     Trace moisture
     Chemical compatibility issues
     Heat tracing requirements
     Other

 

Back To Top ^


 

Model 491M Gas Standards Generator

    The 491M gas standards generator provides:
Industry-standard for over 20 years
Permeation tube oven with wide temperature range controlled to 0.1 °C
Primary dilution mass flow rate
Standby, zero and span modes
Single-oven module can create mixtures with up to 8 components
Expandable

491M features: 

  • Industry standard for over 20 years
  • Complete, integrated, ready-to-use turnkey system
  • Manually operated via front panel controls
  • Direct readout of oven temperature and mass flow rate
  • Modes of Operation: standby, zero and span
  • Designed for expandability 
  • Flow path designed for maximum flexibility with minimum error
  • Flow path suitable for reactive gases - mixture contacts only glass, Teflon® and stainless steel (other materials available)
  • Optional component flow path materials available (sulfinert-coated, all stainless, etc.)
  • Accepts disposable permeation tubes, diffusion tubes, ultra-high rate liquid filled tubes, wafer tubes, and prefilled gas fed permeation tubes
  • Accepts up to 8 disposable tubes per oven with maximum 6 inch length x 1/4 inch diameter (KIN-TEK HRT, SRT and EL tubes)
  • Accepts one of KIN-TEK refillable LFH, ULED and 57 Series tubes per oven
Principle of Operation 491MB
Base Module
     Dilution of constant component vapor emission from permeation or 
     diffusion tubes
     Delivery of precise, adjustable, low concentration dynamically blended
     gas mixtures. Trace concentrations - ppmv, ppbv, pptv
     Manual front panel controls for oven temperature and dilution flow rate
     adjustment and standby, zero and span selection
   
Permeation Tube Oven  
     Construction - high thermal mass oven with electronic PID control Vertical orientation
     Oven temperature range (heat only) 30-150 °C
     NIST-traceable oven temperature
     Oven setpoint resolution and control ±0.1 °C
     Temperature digital indication and adjust via front panel
   
Diluent Gas (customer-supplied)  
     Uses inert, non-reactive, non-hazardous dry gas (Nitrogen, zero air,
     Argon, etc.)
     Mass flow meter subsystem is factory-calibrated in Nitrogen; contact factory
     for calibration in other type gas
       
Primary Dilution Mass Flow Controller
     Dilution flow rate set by manual valve on front panel; measured by mass flow meter
     Standard dilution flow rate range 0.25 to 5 lpm
     Optional flow rate ranges available (0.5-10, 0.1-1, 0.1-0.5 lpm)
     Mass flow rate accuracy ±1% FS
     Mass flow rate indication via front panel
     Mass flow calibration method Analog zero/span adjust
   
Accepts Vapor Emissions from the Following Devices*  
     Liquid-filled disposable tubes up to 15 cm in length (maximum of 8 tubes)
     Liquid-filled high rate (LFH) permeation tubes (maximum of 1 tube)
     Factory pre-filled gas-fed permeation tubes (maximum of 1 tube)
     Diffusion tubes (maximum of 4 tubes)
   
*Note: A single oven can hold only one of the above tube types at a time  
   
Standard Component Flow Path  
     Component contacts only glass, Teflon® and stainless steel
     Glass permeation tube holder
     Suitable for reactive component gases
   
Optional Component Flow Path (specified at time of order)  
     Heated output lines
     Stainless steel
     Electro-polished stainless steel
     Sulfinert-coated stainless steel
     Teflon only
     Teflon permeation tube holder
   
Standard Modes of Operation  
     Standby (oven venting at approx 75 cc/min flow rate)
     Zero (with user-specified flow rate)
     Span-by-flow; flow rate set by manual valve on front panel and measured
     by mass flow meter with front panel digital indication
   
Output Concentration Range  
     Below 1 ppb to over 1000 ppm depending on tube emission rate 
     and dilution flow rate; low ppb and ppt concentrations with secondary dilution
   
Expandability  
     Additional permeation ovens (via up to five 491M-PM modules)
     Customer-refillable gas-fed permeation oven option (via 491M-GF modules)
     Secondary diluition flow rate option (via 491M-SD module); other ranges available

0.5 to 5 lpm

     Overall output gas stream dilution ratio with both primary and secondary dilution Up to 10,000 to 1
     Direct gas blending module option (via 491M-DGB)
     Humidified gas module option (via 491M-HG)
     Gas stream output Interface Module option (via 491M-IM module);
     If 491M-SD secondary dilution module is installed, subtract 10 psig (69 kPa)
ambient to 40 psig (276 kPa)
   
Instrument Housing (specified when ordering)  
     Carrying case (standard)
     Expansion chassis (option); desktop or rack-mount; 2 cases maximum 3 modules per case
     Maximum number of modules in a single 491M system 6
   
Power Requirements (specified when ordering)  
     115 VAC, 50/60 Hz
     230 VAC, 50/60 Hz
   
Dimensions and Weight  
     Dimensions (Base module as packaged in carrying case); approximate 7.5 W x 13.5 H x 20 D inch
18.4 W x 34.3 H x 50.8 D cm
     Weight (Base module as packaged in carrying case); approximate 32 lb (14.5 kg)
   
Certifications  
     European CE Mark (230 VAC, 50 Hz): 491MB (Base), 491M-PM, 491M-SD,
     491M-HG, 491M-GF, 491M-DGB
     UL (site-specific)
   
Speciality Applications (Please contact the Factory)  
     Special diluent (carrier) gases other than inert gases
     Trace moisture
     Chemical compatibility issues
     Heat tracing requirements
     Other

Back To Top ^


 

Span Pac™ H20 Trace Moisture Gas Standard

    The Span Pac™ H20 trace moisture gas standard provides:
H2O concentrations adjustable over 20:1 range with single oven; 400:1 with dual oven system; 10,000:1 with secondary dilution mass flow
Concentrations within 0.01 to 10,000 ppm H2O range depending on model selection
H2O concentrations below 100 ppb with welded model
Industrial models have NEMA 4 cabinets and optional Z or X purge capability
Uses KIN-TEK Trace Source™ disposable and user-refillable water permeation tubes

 

Trace concentration moisture is an extremely important parameter in gases used in a wide range of processes. But there is no absolute analytical method that can be used in all applications. For most sensing methods, the response to moisture is affected by the background gas and most current methods are subject to “poisoning” by certain contaminants. Basically, all trace moisture sensing systems require calibration to external standards and they must be frequently challenged with standards to demonstrate continued accuracy.

There are three special problems associated with trace moisture calibration:

  1. At the lowest concentration levels, there is no real zero reference point for moisture. Standards prepared by simply adding water vapor to a dry gas actually have unknown uncertainty.
     
  2. Water interacts readily with almost all materials, so changes in pressure or temperature can appear as transient concentration changes and may take a very long time to stabilize.
     
  3. Water is pervasive in our atmosphere, so even the most accurate standard may be contaminated in the delivery process.

The Span Pac™ H2O instrument works on a differential principal. Application of this principle is meaningful when the moisture content of the background diluent gas exceeds 10 percent of the minimum concentration to be generated. Calibration is based on the change in analyzer response caused by the addition of a known water vapor concentration from a KIN-TEK Trace Source™ water permeation tube to a flowing stream of the diluent gas. This differential method overcomes the "no zero” problem, and the problem of the background gas affecting the response.

The permeation tube emission rate is certified by the Factory traceable to NIST by gravimetric weight loss over time. This provides a close approximation to a primary standard, even in situations where primary standards are not available. 

KIN-TEK takes special care in design, construction and choice of hardware components to minimize errors due to atmospheric contamination and interactions with other materials in the system. 

Principle of Operation Span Pac™ H20
     Delivery of precise, adjustable concentrations by dilution of moisture vapor
     from one or more certified water permeation tubes held at constant temperature
   
Models Available  
     Span Pac™61-H20 with single oven and primary dilution
     Span Pac™261-H20 with dual ovens and primary dilution
     Span Pac™ 61-H20-SD with single oven and primary and secondary dilution
     Span Pac 61-W-H20 single-oven welded model for concentrations below
     100 ppb (0.01 ppm)
     Span Pac 61-W-H20-SD single-oven welded model with secondary dilution for
     concentrations below 100 ppb (0.01 ppm)
   
Permeation Tube Oven  
     High thermal mass oven with electronic control Horizontal orientation
     Oven temperature range (heat only) 30-150 °C
     NIST-traceable oven temperature
     Oven temperature setpoint and display resolution ±0.1 °C
       
Diluent Gas (customer-supplied)  
     Uses inert, non-reactive, non-hazardous dry gas (nitrogen, zero air, argon, etc.)  
     Mass flow subsystem(s) are factory-calibrated for nitrogen diluent gas. Contact
     Factory for calibration in other diluent gases.
     (option) Instrument can be customized for other diluent gases such as 
     hydrogen or oxygen
Contact Factory
   
Primary Dilution Flow Subsystem  
     Primary dilution mass flow range 0.25 to 5 lpm
     Mass flow rate accuracy ±1% FS
     Mass flow rate indicated on front panel
     Mass flow meter subsystem is factory-calibrated in nitrogen; contact factory
     for calibration in other gases
     (option) 0.5 to 10 lpm; other flow ranges available Contact Factory
   
Secondary Dilution Mass Flow Subsystem (for Span Pac 61-H20-SD model)  
     Secondary dilution mass flow range 0.25 to 5 lpm
     Secondary component split flow mass flow range 10 to 200 sccm
     Mass flow rate accuracy (dilution and component flow controllers) ±1% FS
     Mass flow rate indicated on front panel
     Overall output gas stream dilution ratio with both primary and secondary dilution 10,000:1
     (option) 5 to 100 sccm mass flow range Contact Factory
   
Accepts Vapor Emissions from the Following Devices*  
     Disposable 0.25 in (0.635 cm) water permeation tube(s) up to 5 in (12.7 cm)
     in overall length (maximum of 6 disposable tubes)
     User-refillable liquid filled high rate (LFH) water permeation tube (max. 1 tube)
   
*Note: A single oven can hold only one of the above tube types at a time  
   
Component Flow Path  
     Electro-polished and passivated stainless steel tubing to minimize adsorption
     Electro-polished stainless steel permeation tube holder (for disposable-type tubes)
     Heated lines
   
Cabinet Purge Port  
     Reduces outboard-to-inboard atmospheric moisture contamination
     Maintains an acceptable nitrogen level within the cabinet when using hydrogen
     as the diluent gas to eliminate any possibility of an explosive mixture. (The user
     must supply a continuous controlled flow of nitrogen to the purge port.)
   
Optional Welded Models  
     Seamless tubing with orbitally-welded joints for Span Pac 61-W-H20 and
     Span Pac 61-W-H20-SD models that generate 1 or more concentrations
     below 100 ppb
   
Modes of Operation  
     Standby, zero and span modes set from the front panel
   
Output Concentration Range (depending on permeation tube emission rates)  
     Single oven model: 20 to1 concentration range within 0.01 to 10,000 ppm
     Dual oven model: 400 to 1 concentration range within 0.01 to 10,000 ppm
     Single oven model with secondary dilution: 10,000 to 1 concentration range
     within bounds of 0.01 to 10,000 ppm
     (option) Welded models for concentrations below 100 ppb (0.01 ppm)
     (standard) Output gas stream can be pressurized from ambient to 40 psig for
     single and dual oven models and 35 psig for secondary dilution model
     (option) Other output pressure ranges available Contact Factory
   
Housing Configurations  
     (standard) Desktop enclosure
     (option) NEMA 4 enclosure for industrial environments
     (option) NEMA 4 enclosure with Z-purge to meet NEC Class 1, Group C or D, Division II
     (option) NEMA 4 enclosure with X-purge to adapt unit to Division I requirement
   
Power Requirements (specify when ordering)  
     115 VAC, 50/60 Hz
     230 VAC, 50, 60 or 50/60 Hz
   
Typical Custom Applications (Please contact the Factory)  
     Using trace moisture with reactive gases
     Using other dilution gases such as oxygen, hydrogen, CO2, etc.
     Pressurized output gas stream greater than standard pressures
     Other mass flow rate ranges
     Your special requirements - contact us

Back To Top ^


 

AutoBlend™ Gas Standards Generator

    The AutoBlend™ gas standards generator provides:
Six independent permeation tube ovens with wide temperature range controlled to 0.1 °C
Primary dilution mass flow rate to 10 lpm
Secondary dilution mass flow rate to 10 lpm
Can create mixtures with up to 48 permeation tube components
Dual headspace and humidification options available

 

The AutoBlend™ is an integrated computer-controlled gas blending system designed to create flowing, adjustable-concentration complex gas mixtures and test atmospheres. The system introduces precisely controlled trace concentration components into a flowing, high purity matrix gas.

The AutoBlend uses KIN-TEK Trace Source™ disposable, refillable and diffusion permeation tubes for over 500 chemicals.

The AutoBlend has six permeation tube ovens. Each oven can generate its component concentrations independent of the other ovens.  This is an extremely powerful feature for dynamically testing the effects of a wide range of vapor contaminants since each oven can be switched in (span) and out (vent), and the concentrations generated by each spanning oven are independent of the concentrations generated by the other spanning ovens.

Hardware is included to assure constant gas stream output conditions independent of variations in flow and pressure. This is particulary important for applications where flow capacity is limited (e.g., filling of gas bags), or when back pressure is required to introduce the output gas stream sample to the device under test at an elevated pressure. Controls are provided on the front panel for adjusting the output pressure.
The base gas mixture created in each oven can contain up to eight different analyte components so that a test atmosphere can be spiked with up to 48 components simultaneously.

A small flow of pure dilution gas passes over the permeation (or diffusion) tubes installed in each oven to form a base gas mixture.  A variable portion of each base gas mixture is then added to a main dilution flow to create a primary mixture with variable concentration ratios. 

A secondary dilution feature provides a large dynamic concentration range.  Secondary dilution is used or bypassed automatically in the system as required to produce target concentrations.

In many applications normally innocuous contaminants that occur routinely, or randomly, may be significant for testing purposes.  In the AutoBlend™, two optional dynamic headspace saturation modules can be added (as shown in bottom cabinet above) to allow the addition of vapors from potential transient interfering compounds such as cleaning solutions, fuels or paints.  Output from the headspace modules can be added either before or after secondary dilution of the trace components.

The resulting test atmosphere can be supplied either dry or at variable humidity levels with the optional humidification module.  Humidity is added by splitting the total dilution flow into two parts.  One part remains dry and contains all of the trace components from the permeation tubes and headspace modules.  The other portion of the dilution gas is saturated with water vapor.  The two portions are combined at the output of the system to form the humidified mixture.  Relative humidity (RH) is varied by varying the ratio of humidified diluent to dry diluent. The trace components do not contact liquid water in the system.

Principle of Operation AutoBlend™
     Delivery of precise, adjustable, low concentration dynamically blended
     gas mixtures. Trace concentrations - ppmv, ppbv, pptv
     Primary dilution of constant component vapor emission from permeation or 
     diffusion tubes. Secondary dilution is automatically added but only if needed to
     bring the target concentrations to specified values
     Method-driven. Up to 20 points of multi-component concentration mixtures
     can be added to a user-named method.
   
Permeation Tube Oven  
     Six independent high thermal mass ovens with electronic PID control Vertical orientation
     Oven temperature range (heat only) 30-150 °C
     NIST-traceable oven temperature
     Oven setpoint resolution and control ±0.1 °C
     Split flow mass flow controller with each oven 10 to 200 sccm
     Independent oven over-temperature alarms
   
Diluent Gas (customer-supplied)  
     Uses inert, non-reactive, non-hazardous dry gas (Nitrogen, zero air,
     Argon, etc.)
     Mass flow meter subsystem is factory-calibrated in Nitrogen; contact factory
     for calibration in other type gas
       
Primary Dilution Mass Flow Subsystem  
     Primary dilution flow rate auto-set in software as a function of target concentration
     Primary dilution mass flow controller range 0.5 to 10 lpm
     Mass flow rate accuracy ±1% FS
     Mass flow rate indication via PC computer screen (with AutoBlend software)
   
Secondary Dilution Mass Flow Subsystem  
     Secondary dilution flow rate auto-set in software as a function of target concentration
     Secondary dilution mass flow controller range 0.5 to 10 lpm
     Secondary component split flow mass flow controller range 10 to 200 sccm
     Mass flow rate accuracy (dilution and component flow controllers) ±1% FS
     Mass flow rate indication via PC computer screen (with AutoBlend software)
     Overall output gas stream dilution ratio with both primary and secondary dilution 10,000:1
   
Accepts Vapor Emissions from the Following Devices*  
     Liquid-filled disposable tubes up to 15 cm in length (maximum of 8 tubes)
     Liquid-filled high rate (LFH) permeation tubes (maximum of 1 tube)
     Factory pre-filled gas-fed permeation tubes (maximum of 1 tube)
     Diffusion tubes (maximum of 4 tubes)
     Headspace (for up to two optionally installed headspace modules only)
   
*Note: A single oven can hold only one of the above tube types at a time  
   
Standard Component Flow Path  
     Component contacts only glass, Teflon® and stainless steel
     Glass permeation tube holder
     Suitable for reactive component gases
   
Optional Component Flow Path (specified at time of order)  
     Heated output lines
     Stainless steel
     Electro-polished stainless steel
     Sulfinert-coated stainless steel
     Teflon only
     Teflon permeation tube holder
   
Standard Modes of Operation  
     Standby (oven venting at approx 75 to 100 cc/min flow rate)
     Zero (with user-specified flow rate)
     Span-by-concentration. Oven and headspace module split flow controllers,
     primary and secondary dilution flow controllers and the secondary component
     split flow controller are automatically set by the software in order to achieve
     the target concentrations.
   
Output Concentration Range  
     Below 1 ppb to over 1000 ppm depending on tube emission rate 
     and primary and secondary dilution flow rates
   
Expandability (specify options when ordering)  
     Dual headspace module option (each with 200 sccm split mass flow controller)
     Humidified gas module option (with 10 lpm mass flow controller)
   
Power Requirements (specified when ordering)  
     115 VAC, 50/60 Hz
     230 VAC, 50/60 Hz
   
Speciality Applications (Please contact the Factory)  
     Special diluent (carrier) gases other than inert gases
     Trace moisture
     Chemical compatibility issues
     Heat tracing requirements
     Other

Back To Top ^