Isco 3000 Series Flow Transmitters are AC powered for permanent installation. They combine dependable flow measurement with versatile data transmission, to give you economical flow monitoring with the features and reliability you demand. Field wiring is connected through conduit, or with cables and sealed compression fittings.
Fixed-site flow/level transmitter with non-contacting Ultrasonic sensor.
3010 Ultrasonic Flow Transmitter
Fixed-site flow/level transmitter with non-contacting Ultrasonic sensor.
The 3010 is AC powered for permanent installation. It has no internal data storage, but provides 4-20 mA analog and RS-232 serial outputs, as well as relay switches and pulsed DC totalizer outputs for flexible interfacing and control. Standard Features
Applications
Options and Accessories
|
Specifications
Isco 3010 Ultrasonic Flow Transmitter |
|
Size (HxWxD): | 15.3 x 10.6 x 7.38 in (38.7 x 27.0 x 18.7 cm) |
Weight: | 10 lbs (4.5 kg) |
Material: | High-impact molded polystyrene structural foam |
Enclosure: | Nema 4X (IP56) |
Power: | 104 to 127V AC, 0.075 A; or 208 to 254V AC, 0.038 A, 50/60 Hz |
Operating Temperature: | -20° to 140°F (-30° to 60°C) |
Storage Temperature: | -50° to 150°F (-46° to 66°C) |
Display: | 6-digit backlit LCD, digits 0.5 inch (1.27 cm) high |
Totalizer: | 7 digit mechanical, non-resettable |
Level-to-Flow Rate Conversions |
|
WEIRS: | V-notch, rectangular with and without end contractions, Cipolletti. |
FLUMES: | Parshall, Palmer-Bowlus, Leopold-Lagco, trapezoidal, H. |
MANNING FORMULA: | Round, U-channel, rectangular, trapezoidal. |
EQUATIONS: | Two-term polynomial of the form K^1H^N1 ±K^2H^N2. |
Characterization PROM: | (optional) Factory programmed with any level-to-flow rate relationship |
Communications and I/O |
|
Sampler pacing output: | Isolated contact closure, rated 1 A at 48V DC |
Sampler input: | Event mark, bottle number |
Analog output: | Isolated 4 to 20 mA based on level or flow rate, with or without sampler event marks, into a maximum of 1000 ohms |
Relay outputs: | 2 form C relays with field selectable trip points based on flow rate, rated 3 A at 24V AC or DC (with optional High/Low Alarm Relays) |
Serial data output: | RS-232 ASCII string of level, flow rate, total flow, sample number and bottle number, once every 12 seconds |
Format: | 300 baud, 7 data bits, 2 stop bits, even parity |
Remote totalizer output: | 12V pulse, compatible with Isco Remote Totalizer |
Ultrasonic Sensor |
|
Length x diameter: | 6.9 x 3.6 in (17.5 x 9.1 cm) |
Cable: | 25 ft x 0.3 in (7.6 m x 0.8 cm) |
Weight (incl cable): | 2.6 lbs (1.2 kg) |
Enclosure: | NEMA 4X, 6; IP67 |
Operating temperature: | -22° to 140°F (-30° to 60°C) |
Compensated temperature: | -22° to 140°F (-30° to 60°C) |
Materials: | Sensor housing is Xenoy 6120; Cable jacket is Polyvinyl chloride (PVC) |
Frequency: | 40 kHz |
Range (distance from sensor to liquid): | Minimum 2 ft (0.6 m); Maximum 12 ft (3.7 m) |
Span: | 0 to 10 ft (0 to 3 m) |
Blanking distance: | 2 to 12 ft. (0.6 to 3.7 m) |
Level Measurement Accuracy - at 22°C (72°F), still air, 40 to 70% humidity. Head change is actual change in vertical distance from ultrasonic sensor to liquid surface) |
|
for head change <1.0 ft (<0.31 m): | maximum error is ±0.02 ft. (±0.006 m) |
for head change of 1.0 to 10 ft (0.31 to 3.05 m): | maximum error is ±0.03 ft. (±0.009 m) |
Temperature coefficient: | Maximum error over compensated temperature range (per degree of temperature change) ±0.000047 x D per °F (±0.000085 x D per °C) where D is the distance from the transducer to the liquid surface |